Fluke 1529 Chub-E4 Standards Thermometer
Precision temperature measurement for professional calibration and testing
Fluke 1529 Chub-E4 Standards Thermometer is a professional precision measurement solution, making Fluke 1529 Chub-E4 Standards Thermometer Price in Bangladesh an important consideration for calibration laboratories, engineers, metrology professionals, and industrial technicians. Designed for demanding temperature measurement applications, this standards thermometer supports reliable monitoring and calibration activities where measurement stability and accuracy are essential. Its professional construction makes it suitable for laboratory, industrial, research, and quality control environments.
The Fluke 1529 Chub-E4 is designed to support precision temperature measurement using compatible reference sensors. Reliable temperature measurement is essential in calibration and metrology because even small measurement variations can affect testing and quality assurance results. The instrument provides a professional platform for users who require dependable temperature data during calibration, verification, and controlled measurement processes.
A precision standards thermometer is an important instrument for laboratories and organizations responsible for maintaining measurement accuracy. The Fluke 1529 Chub-E4 can support calibration workflows by providing a dependable reference measurement platform. It is suitable for applications where temperature instruments need to be evaluated, verified, or compared against a trusted measurement reference.
The Fluke 1529 Chub-E4 is well suited to laboratory environments where controlled and repeatable temperature measurement is required. It can assist laboratory technicians with temperature verification, equipment calibration, quality assurance, and research activities. Its professional design allows it to fit into established measurement procedures while supporting the requirements of technical users.
Industrial processes often require dependable temperature measurement for equipment inspection, process control, and maintenance. The Fluke 1529 Chub-E4 can be used with suitable sensors for professional temperature monitoring and verification tasks. Engineers and maintenance technicians can incorporate the instrument into procedures where reliable reference temperature measurements are important for evaluating equipment and process conditions.
The instrument is designed for users who require a higher level of measurement performance than conventional temperature monitoring equipment. When paired with an appropriate precision sensor, the Fluke 1529 Chub-E4 can provide a dependable measurement solution for professional calibration and testing. Proper sensor selection and calibration practices are essential for achieving the required performance in any specific application.
Professional measurement equipment must remain dependable during regular laboratory and field use. The Fluke 1529 Chub-E4 features a purpose built design intended for precision temperature measurement applications. Careful handling, proper sensor connection, suitable operating conditions, and routine verification can help maintain reliable performance throughout the instrument's service life.
This standards thermometer is suitable for calibration engineers, laboratory technicians, metrology specialists, quality control professionals, researchers, and industrial maintenance teams. It can support a variety of measurement activities where temperature accuracy and repeatability are important. Its professional application range makes it a valuable component of a well managed temperature calibration system.
The Fluke 1529 Chub-E4 Standards Thermometer can support calibration laboratories, manufacturing facilities, pharmaceutical operations, research institutions, industrial service providers, and quality assurance departments in Bangladesh. It can be used for temperature verification, calibration support, laboratory testing, equipment assessment, and technical measurement tasks. Users should confirm sensor compatibility and application requirements before integrating the instrument into an existing measurement system.
The Fluke 1529 Chub-E4 combines precision measurement capability with a professional design intended for demanding calibration and laboratory applications. Its suitability for technical environments makes it a practical choice for organizations that depend on consistent and trustworthy temperature measurements. With compatible reference sensors and proper calibration procedures, it can contribute to reliable temperature measurement and quality assurance programs.
| PRT / RTD | Thermistor | Thermocouple | |
| Inputs | 2 channels PRT/thermistor and 2 channels TC, or 4 channels PRT/thermistor, or 4 channels TC, specify when ordering; PRT/thermistor channels accept 2, 3, or 4 wires; TC inputs accept B, E, J, K, N, R, S, T, and Au-Pt TC types. (Support for C and U type thermocouples is available. Download the application note Using Fluke Calibration Readouts with Tungsten-Rhenium and other Thermocouples) |
||
| Temperature Range | –189 °C to 960 °C | –50 °C to 150 °C | –270 °C to 1800 °C |
| Measurement Range | 0 to 400 Ω | 0 to 500 KΩ | –10 to 100 mV |
| Characterizations | ITS-90, IEC-751 (DIN “385”), Callendar-Van Dusen | Steinhart-Hart, YSI-400 | NIST Monograph 175, 3-point deviation function applied to NIST 175, 6th-order polynomial |
| Temperature Accuracy (meter only) | ± 0.004 °C at –100 °C ± 0.006 °C at 0 °C ± 0.009 °C at 100 °C ± 0.012 °C at 200 °C ± 0.018 °C at 400 °C ± 0.024 °C at 600 °C |
± 0.0025 °C at 0 °C ± 0.0025 °C at 25 °C ± 0.004 °C at 50 °C ± 0.010 °C at 75 °C ± 0.025 °C at 100 °C |
Ext. RJC Int. RJC B at 1000 °C ± 0.6 °C ± 0.6 °C E at 600 °C ± 0.07 °C ± 0.25 °C J at 600 °C ± 0.1 °C ± 0.35 °C K at 600 °C ± 0.15 °C ± 0.4 °C N at 600 °C ± 0.15 °C ± 0.3 °C R at 1000 °C ± 0.4 °C ± 0.5 °C S at 1000 °C ± 0.5 °C ± 0.6 °C T at 200 °C ± 0.1 °C ± 0.3 °C |
| Temperature Resolution | 0.001 ° | 0.0001 ° | 0.01 to 0.001 ° |
| Resistance/Voltage Accuracy | 0 Ω to 20 Ω: ± 0.0005Ω 20 Ω to 400 Ω: ± 25 ppm of rdg. |
0 Ω to 5 KΩ: ± 0.5 Ω 5 KΩ to 200 KΩ: ± 100 ppm of rdg. 200 KΩ to 500 KΩ ± 300 ppm of rdg. |
–10 to 50 mV: ± 0.005 mV 50 to 100 mV: ± 100 ppm of rdg. Internal RJC: ± 0.25 °C) |
| Operating Range | 16 °C to 30 °C | ||
| Measurement Interval | 0.1 second to 1 hour; inputs may be read sequentially or simultaneously at 1 second or greater interval | ||
| Excitation Current | 1 mA, reversing | 2 and 10 µA, automatically selected | n/a |
| Display | 33 x 127 mm (1.3 x 5 in) backlit LCD graphical display | ||
| Display Units | °C, °F, K, Ω, KΩ, mV | ||
| Data Logging | Up to 8,000 time- and date-stamped measurements can be logged | ||
| Logging Intervals | 0.1, 0.2, 0.5, 1, 2, 5, 10, 30, or 60 seconds; 2, 5, 10, 30, or 60 minutes | ||
| Averaging | Moving average of most recent 2 to 10 readings, user selectable | ||
| Probe Connection | Patented DWF Connectors accept mini spade lug, bare-wire, or mini banana plug terminations | Universal receptacle accepts miniature and standard TC connectors | |
| Communications | RS-232 included, IEEE-488 (GPIB) optional | ||
| AC Power | 100–240 V ac, 50-60 Hz, 0.4 A | ||
| DC Power | 12–16 VDC, 0.5 A (battery charges during operation from 14.5 to 16V DC, 1.0A) | ||
| Battery | NiMH, 8 hours of operation typical without backlight, 3 hours to charge, 500 cycles | ||
| Size (HxWxD) | 102 x 191 x 208 mm (4.0 x 7.5 x 8.2 in) | ||
| Weight | 2 kg (4.5 lb) | ||
| Calibration | Accredited resistance and voltage calibrations are provided. | ||
Cases and holsters (1)
Rack mounts and hardware (1)
No resources available for this product.
No reviews yet. Be the first to review this product!