Fluke 7340, 7380 Compact Temperature Calibration Baths
Compact thermal performance for precision temperature calibration
Fluke 7340 and 7380 are compact temperature calibration baths designed for controlled laboratory temperature applications, making Fluke 7340, 7380 Compact Temperature Calibration Baths Price in Bangladesh a relevant search term for calibration professionals and industrial laboratories. These baths provide a stable liquid temperature environment for calibrating temperature sensors and instruments where repeatability, thermal stability, and dependable measurement performance are important.
The compact design of these calibration baths makes them practical for laboratories where workspace and equipment mobility are important considerations. Their focused form factor allows engineers to integrate the bath into existing calibration benches without requiring excessive installation space. This makes them suitable for both dedicated calibration environments and industrial testing facilities.
Temperature stability is essential for reliable calibration work. The Fluke compact baths are designed to establish a controlled thermal environment that helps maintain consistent conditions around the sensors being tested. Stable liquid temperature distribution supports repeatable measurements and helps calibration technicians obtain dependable results during routine and advanced testing procedures.
A temperature calibration bath works by immersing the sensor or device under test in a controlled liquid medium. This provides a uniform thermal environment around the sensing element and helps minimize temperature gradients. The Fluke 7340 and 7380 are suited to applications where consistent thermal conditions are required for comparing reference sensors with devices under test.
These calibration baths can be used with a range of temperature sensing devices suitable for liquid bath calibration. They can support calibration activities involving resistance temperature detectors, thermocouples, temperature probes, and other compatible sensors. Their controlled thermal environment makes them useful for laboratories performing verification and calibration procedures.
The baths are suitable for calibration laboratories, industrial maintenance departments, quality control facilities, research organizations, and technical service providers. They can be integrated into workflows where temperature instruments need to be tested against a known and controlled thermal condition. This helps organizations maintain consistent measurement practices across different applications.
Repeatability is a key requirement in professional temperature calibration. Consistent bath operation helps technicians perform repeated tests under comparable thermal conditions. This can improve confidence in calibration results and make the equipment suitable for regular verification programs where reliable temperature measurement is important.
The Fluke baths can form part of a broader temperature calibration system that includes reference thermometers, digital measurement instruments, sensor accessories, and calibration documentation. Their compact design allows them to be positioned conveniently within laboratory workstations and technical service environments, supporting efficient calibration workflows.
Professional calibration equipment must provide dependable performance during repeated use. The Fluke compact temperature baths are designed for laboratory environments where controlled thermal operation is essential. Proper handling, suitable operating procedures, and routine maintenance can help preserve thermal performance and support long term equipment reliability.
Calibration laboratories, manufacturing facilities, pharmaceutical organizations, food processing industries, research institutions, and technical service companies in Bangladesh may require reliable temperature calibration equipment. The Fluke 7340 and 7380 can support organizations that need compact and practical thermal equipment for temperature sensor verification and calibration activities.
When choosing a compact temperature calibration bath, engineers should evaluate temperature stability, uniformity, liquid compatibility, sensor size, laboratory requirements, operating environment, and service support. Proper selection ensures that the bath can accommodate the intended calibration workload and work effectively with existing measurement equipment.
Regular maintenance is important for preserving the performance of a temperature calibration bath. Appropriate bath fluid selection, cleanliness, proper sensor placement, and routine inspection can help maintain stable operation. Calibration checks and professional servicing can further support measurement reliability throughout the equipment's service life.
The Fluke 7340, 7380 Compact Temperature Calibration Baths provide a practical solution for laboratories requiring controlled liquid temperature environments. Their compact design, stable thermal performance, and suitability for professional temperature sensor calibration make them valuable tools for industrial and laboratory applications. For organizations in Bangladesh seeking dependable temperature calibration equipment, these baths can support efficient, repeatable, and well controlled measurement procedures.
| Specifications | ||
| 7340 | 7380 | |
| Range | –40 °C to 150 °C | –80 °C to 100 °C |
| Stability | ± 0.005 °C at –40 °C (ethanol) ± 0.005 °C at 25 °C (water) ± 0.007 °C at 150 °C (oil 5012) |
± 0.006 °C at –80 °C (ethanol) ± 0.010 °C at 0 °C (ethanol) ± 0.010 °C at 100 °C (oil 5012) |
| Uniformity | ±0.006 °C at –40 °C (ethanol) ±0.005 °C at 2 5 °C (water) ±0.010 °C at 150 °C (oil 5012) |
±0.008 °C at –80 °C (ethanol) ±0.012 °C at 0 °C (ethanol) ±0.012 °C at 100 °C (oil 5012) |
| Heating Time† | 60 minutes, from 25°C to 150°C (oil 5012) |
25 minutes, from 25 °C to 100 °C (oil 5010) |
| Cooling Time | 110 minutes, from 25°C to –40°C (ethanol) |
130 minutes, from 25 °C to –80 °C (ethanol) |
| Stabilization Time | 15–20 minutes | |
| Temperature Setting | Digital display with push-button data entry | |
| Set-Point Resolution | 0.01°; 0.00018° in high-resolution mode | 0.01° |
| Display Resolution | 0.01 ° | |
| Digital Setting Accuracy | ± 0.5 °C | |
| Digital Setting Repeatability | ± 0.01 °C | |
| Access Opening | 94 x 172 mm (3.7 x 6.8 in) | 86 x 114 mm (3.25 x 4.5 in) |
| Working Area | 81 x 133 mm (3.2 x 5.25 in) | 86 x 114 mm (3 x 4 in) |
| Depth | 234 mm (9.25 in) | 178 mm (7 in) |
| Wetted Parts | 304 stainless steel | |
| Power | 115 V ac (±10 %), 60 Hz, 15 A or 230 V ac (±10 %), 50 Hz, 8 A, specify, 1400 VA |
115 V ac (±10 %) 60 Hz, 16 A or 230 V ac (±10 %),50 Hz, 8 A, specify |
| Volume | 9.2 liters (2.4 gal) | 4 liters (1 gal) |
| Size (WxDxH) | 305 x 622 x 584 mm (12 x 24.5 x 23 in) off cart; 305 x 622 x 819 mm (12 x 24.5 x 32.25 in) on cart | 305 x 610 x 762 mm (12 x 24 x 30 in) |
| Weight | 35.4 kg (78 lb) | 52 kg (115 lb) |
| Automation | Use the 1586A Super-DAQ to automate temperature sensor calibration | |
| †Rated at nominal 115 V (or optional 230 V) | ||
N/A
No resources available for this product.