Fluke 789 Process Meter
Precision Process Measurement and Calibration Made Easier
The Fluke 789 Process Meter is a versatile instrument for industrial electrical testing and process calibration, making the Fluke 789 Process Meter Price in Bangladesh an important consideration for engineers and technicians seeking a dependable solution for process measurement and troubleshooting. Its multifunctional design combines digital multimeter capabilities with process current sourcing and simulation functions, helping professionals handle a wide range of instrumentation tasks with greater efficiency.
The Fluke 789 Process Meter is designed to support accurate electrical measurement and process signal testing in demanding industrial environments. It enables technicians to evaluate current signals and perform essential diagnostic work on process control systems. This makes it a practical instrument for maintenance, commissioning, and troubleshooting activities.
A key advantage of the Fluke 789 is its ability to source and simulate current signals for process applications. These functions are useful when testing transmitters, control loops, indicators, and other instrumentation equipment. Engineers can verify signal response and identify problems without relying on several separate testing instruments.
Industrial process systems often depend on reliable current loops for transmitting measurement and control signals. The Fluke 789 Process Meter helps technicians investigate loop-related issues efficiently by providing measurement, sourcing, and simulation capabilities in one instrument. This can simplify field diagnostics and reduce the time required to identify signal problems.
Designed with professional fieldwork in mind, the Process Meter offers a practical combination of functionality and usability. Its controls are arranged to provide convenient access to frequently used measurement and calibration functions. The rugged construction makes it suitable for engineers and maintenance personnel working in industrial and process environments.
The Fluke 789 Process Meter can be used in manufacturing facilities, process plants, industrial automation systems, electrical maintenance departments, instrumentation workshops, and control system commissioning projects. It is particularly useful for professionals involved in transmitter testing, loop verification, process calibration, and electrical troubleshooting.
Using a multifunctional process meter can help maintenance teams streamline their workflow. Instead of switching between separate instruments for different process tasks, technicians can perform several common measurements and signal tests with one professional tool. This contributes to more organized troubleshooting and efficient maintenance operations.
The Fluke 789 Process Meter is well suited to instrumentation engineers, electrical technicians, automation professionals, and industrial maintenance teams who require dependable testing equipment. Its combination of electrical measurement and process calibration capabilities makes it a valuable tool for maintaining and commissioning modern industrial control systems.
For professionals searching for a reliable process measurement and troubleshooting instrument, the loop calibrator offers a practical combination of versatility, durability, and engineering-focused functionality. Its ability to measure, source, and simulate process signals makes it suitable for demanding calibration and maintenance applications where dependable field performance is essential.
| Specifications | ||
| Current measurement | Range | 0-1 A AC or DC |
| Resolution | 1 mA | |
| Accuracy | 0.2% + 2 LSD (DC) | |
| Range | 0-30 mA | |
| Resolution | 0.001 mA | |
| Accuracy | 0.05% + 2 lSD | |
| Current sourcing | Range | 0-20 mA or 4-20 mA |
| Resolution | 0.05% of span | |
| Drive capability | 24 V Compliance or 1200 Ω @ 20 mA | |
| Frequency | to 19.999 kHz, 0.005% + 1 LSD | |
| Continuity | Beeps for resistance < 100 Ω | |
| Voltage measurement | Range | 0-1000 V AC (true-rms) or DC |
| Resolution | 0.1 mV to 1.0 V | |
| Accuracy | 0.1% Rdg + 1 LSD (VDC) | |
| Resistance measurement | to 40 MΩ, 0.2% + 1 LSD | |
| Diode test | 2.0 V shoes diode voltage drop | |
| Display current and % of span | Yes | |
| Auto Step, Auto Ramp | Yes | |
| Environmental Specifications | ||
| Operating temperature | -20°C to 55°C | |
| Storage temperature | -40°C to 60°C | |
| Humidity (without condensation) | 95% up to 30°C 75% up to 40°C 45% up to 50°C 35% up to 55°C |
|
| Temperature coefficient | 0.05 x (specified accuracy) per °C (for temperatures < 18°C or > 28°C) | |
| Safety Specifications | ||
| Electrical safety | Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No 1010.1-92 | |
| Overvoltage category | IEC 1010-1 CAT III 1000 V IEC 1010-1 CAT IV 600 V |
|
| Maximum voltage | 1000 V RMS | |
| Mechanical and General Specifications | ||
| Size (H x W x L) | 50 x 100 x 203 mm (1.25 x 3.41 x 7.35 in) | |
| Weight | 610 g (1.6 lbs) | |
| Batteries | 4 AA alkaline batteries | |
| Warranty | 3 years | |
| Battery life | 140 hours typical (measurement), 10 hours typical (sourcing 20 mA) | |
| Shock and vibration | Random, 2 g, 5-500 Hz 1 meter drop test |
|
No resources available for this product.